California Instruments 801RP Series Spezifikationen Seite 92

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 126
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 91
10.2.3 Output Calibration
The AC voltage output calibration is performed by adjusting the output up or down using a
calibration coefficient. The coefficient ranges from -127 (adjust upward) to + 127 (adjust
downward). An external traceable reference AC voltmeter should be used for this purpose.
CALibrate[:SOURce]:VOLTage <NRf>
This command will set the calibration coefficient for the AC full scale output voltage.
Command Syntax CALibrate[:SOURce]:VOLTage <NRf>
Parameters <NRf> (a value between -127 and +127)
Examples CAL:VOLT -2
Query Syntax CALibrate:VOLTage?
Returned Parameters <NR1> (value range -127 to +127)
Related Commands CAL:STAT
Seitenansicht 91
1 2 ... 87 88 89 90 91 92 93 94 95 96 97 ... 125 126

Kommentare zu diesen Handbüchern

Keine Kommentare