California Instruments 801RP Series Spezifikationen Seite 91

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 126
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 90
CALibrate:MEASure:POWer[:FSCale] <NRf>
This command affects the calibration of the power measurement at full scale.
Command Syntax CALibrate:MEASure:POWer[:FSCale] <NRf>
Parameters <NRf> (value range -127 to +127)
Examples CAL:MEAS:POW 78
Query Syntax CALibrate:MEASure:POWer?
Returned Parameters <NR1> (value range -127 to +127)
Related Commands CALibrate:MEASure:CURRent
CALibrate:MEASure:POWer:ZERO <NRf>
This command affects the calibration of the power measurement zero offset.
Command Syntax CALibrate:MEASure:POWer:ZERO <NRf>
Parameters <NRf> (value range -127 to +127)
Examples CAL:MEAS:POW:ZERO -5
Query Syntax CALibrate:MEASure:POWer:ZERO?
Returned Parameters <NR1> (value range -127 to +127)
Related Commands CALibrate:MEASure:POWer
CALibrate:MEASure:VOLTage[:FSCale] <NRf>
This command affects the calibration of the rms voltage measurement at full scale.
Command Syntax CALibrate:MEASure:VOLTage[:FSCale] <NRf>
Parameters <NRf> (value range -127 to +127)
Examples CAL:MEAS:VOLT 56
Query Syntax CALibrate:MEASure:VOLTage?
Returned Parameters <NR1> (value range -127 to +127)
Related Commands CALibrate:MEASure:CURRent
CALibrate:MEASure:VOLTage:ZERO <NRf>
This command affects the calibration of the rms voltage measurement zero offset.
Command Syntax CALibrate:MEASure:VOLTage:ZERO <NRf>
Parameters <NRf> (value range -127 to +127)
Examples CAL:MEAS:VOLT:ZERO +4
Query Syntax CALibrate:MEASure:VOLTage:ZERO?
Returned Parameters <NR1> (value range -127 to +127)
Related Commands CALibrate:MEASure:VOLTage
Seitenansicht 90
1 2 ... 86 87 88 89 90 91 92 93 94 95 96 ... 125 126

Kommentare zu diesen Handbüchern

Keine Kommentare