California Instruments 801RP Series Spezifikationen Seite 117

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 126
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 116
To measure the RMS current, use:
MEAS:CURR?
To measure the peak current, use:
MEAS:CURR:AMPL:RES /* resets track and hold for peak curr. meas.
/* It may be need to turn the output on or program
/* as specific voltage at this time before taking a
/* peak current reading.
MEAS:CURR:AMPL:MAX? /* retrieves peak current reading.
To measure the current crest factor, use:
MEAS:CURR:CRES?
Power Measurements
To measure the true power, use:
MEAS:POW:REAL?
To measure the apparent power, use:
MEAS:POW:APP?
To measure the power factor, use:
MEAS:POW:PFAC?