AMETEK Programmable Power MX / RS Series SCPI Programming Manual
170 MX / RS Series
8.2 IEC 1000-4-11 (-411)
The following are the remote commands set for IEC1000-4-11. The ABORt command will
terminate a test in progress at any time. Test completion can be checked using the *OPC?
query if the *OPC 1 command is issued immediately after an IEC1000-4-11 command. Upon
completion of the test run *OPC? will return a 1. This may be relevant for ATE applications
in view of long test times for some IEC 1000-4-11 test sequences.
Details on the actual tests performed as a result of each of these commands may be found
in the IEC 1000-4-11 option section of the User Manual which is supplied with the AC/DC
power source.
IEC411:
[:NOMinal]
:VOLTage
:DIPS
:RUN
[:ALL]
:SINGle
:VOLTage
[:PERCent]
:CYCLes
:ANGLe
:STATe
:VARiants
:RUN
[:ALL]
:SINGle
:HOLD
:TIME
:VOLTage
[:PERCent]
:FALL
[:TIME]
:RISE
[:TIME]
IEC411[:NOMinal]:VOLTage
This command set the nominal voltage for the test
Command Syntax IEC411[:NOMinal]:VOLTage <NRf>
Parameters <volatge>
Examples IEC411:VOLT 230
Query Syntax IEC411:VOLT?
Returned Parameters <NR2>