User Manual 2253i / 2253iX 1. Introduction This instruction manual (P/N 6005-962) contains information on the installation, operation, calibration
User Manual 2253i / 2253iX 7.3.3 Overload Light is On CAUSE SOLUTION Unit is overloaded Remove overload or check CL setting Unit is switched to
User Manual 2253i / 2253iX 7.4 Self test An internal self-test can be performed over the bus by sending the *TST? query command. The self-test will
User Manual 2253i / 2253iX 7.6 Amplifier Module Data This section lists the various connectors and interface pin outs to the 6005-400-1 amplifier
User Manual 2253i / 2253iX 7.6.2 P OWER BOARD7.6.2.1 HIGH VOLTAGE DC INPUT PINS NAME DESCRIPTION E1 +250V_IN ink current from module +250 Vdc
User Manual 2253i / 2253iX DIP SWITCH POSITION NAME FUNCTION different master. 2 IC ON to connect common of module’s LEFT amplifier as the maste
User Manual 2253i / 2253iX 7.9 Repla blcea e Parts In order to ensure prompt, accurate service, please provide the following information, when appl
User Manual 2253i / 2253iX REF # Sub CI PART # DESCRIPTION MNF, P/N QTY A1 6005-700-1 Amplifier Power Board CI 1 F1, F2 270238 Fuse,
User Manual 2253i / 2253iX 8. Miscellaneous Options 8.1 IEEE488 Interface (-GPIB) The GPIB interface is available on all iX Series models. It is a
User Manual 2253i / 2253iX 8.4 External Input (-EXT) The –EXT option is factory installed only. If installed, this option disables the internal c
User Manual 2253i / 2253iX 8.5 RPV Input (-RPV) The –RPV option is factory installed only. If installed, this option disables the internal contro
User Manual 2253i / 2253iX 1.2 iX and i Model Differences The 2253iX offers additional features and functions over the 2253i model. Some of the fe
User Manual 2253i / 2253iX 8.8 Rack Mount Slides (-RMS) For mounmount slidting the 2253i/iX in a 19-inch instrument cabinet, it is recommended to
User Manual 2253i / 2253iX 9. Option -160: RTCA / DO-160 Rev D, E 9.1 General Option –160 includes a firmware implementation for RTCA/DO160 revisi
User Manual 2253i / 2253iX 9.3 Available DO160 Tests 9.3.1 NORMAL STATE AC Mst 2. Voltage unbalance test 3. Waveform Distortion test 4. Voltage
User Manual 2253i / 2253iX 5. Voltage Surge 9.4 Front Panel Operation -160 To perform a test froAPPLICATIONS/OPm the keyboard, Press the MENU key
User Manual 2253i / 2253iX 9.5 AC Test Mode F ov testing utput mod9.5.1 Normal State tests ollow ions cing sect er in o AC e. Scroll to the NO
User Manual 2253i / 2253iX VOLT FREQ MIN Standard/Group RTCA A(CF) A(NF) A(WF) 1 100 00 00 0 1 1 10Voltage 3 1.5 5 10 101. 101.5 101.5
User Manual 2253i / 2253iX VOLT UNBALANCE Standard/Group RTCA A(CF) A(NF) A(WF) Voltage offset 6 6 6 8 Frequency 400 390/410 360/650 360/80
User Manual 2253i / 2253iX Figure 9-4: Voltage Modulation - Frequency characteristics California Instruments 117
User Manual 2253i / 2253iX FREQUENCY MODULATION This test requires a numeric value equal to the modulation rate in Hz. This value must be between
User Manual 2253i / 2253iX POWER INTERRUPT lue equal to the test number. The tests are grouped as follows: s 1 through 15 are for all Standard and
User Manual 2253i / 2253iX 2. Specifications Specifications shown are valid over an ambient temperature range of 25 ± 5° C and apply after a 30 min
User Manual 2253i / 2253iX T1 VolF1 F2 T2 0t T3 Test no.: 21 22 23 24 25 26 Standard: I II
User Manual 2253i / 2253iX FREQUENCY TRANSIENTS (Group 1 and A(CF) only) Seq. No Frequency Time 1 400 5 Minute 2 440 150msec 3 420 1.5sec 4 400
User Manual 2253i / 2253iX 9.5.2 EMERGENCY TEST From the DO160 MENU scroll to the EMERGENCY AC entry using the up and down cursor keys. Press the E
User Manual 2253i / 2253iX VOLT FREQ MAX This test will set the voltage and frequency to a level defined by Table 9-9. The test will last for 30 VO
User Manual 2253i / 2253iX 9.5.3 ABNORMAL TEST From the DO160 MENU Scroll to the ABNORMAL AC entry using the up and down cursor keys. Presthe ENTER
User Manual 2253i / 2253iX VOLT MIN This test will set the voltage and frequency to levels defined by Table 9-12 for 5 minutes. The test will be d
User Manual 2253i / 2253iX FREQUENCY TRANSIENTS (A(CF) only) Test 1 Seq. No. Volt/Frequency Time 1 115/400 5 minutes 2 115/350 5 sec. 3 115/320
User Manual 2253i / 2253iX 9.6 DC Test Mode DC test mode requires DC mode and a steady state voltage setting of 24V DC or 14V DC. Note: Prior to
User Manual 2253i / 2253iX VOLT MAX Standard Categories A and Z B 28V / 14V RTCA 30.3 30.3 15.1 EUROCAE 30.3 29.3 14.6 Table 9-16: Normal Volta
User Manual 2253i / 2253iX 9.6.2 Abnormal Test From e 9-11. the DO-160 MENU scroll to the ABNORMAL DC entry using the up and down cursor keys. Pr
User Manual 2253i / 2253iX 2.1.2 Output Output Parameter Specification Model 2253i 2253iX Modes AC, DC AC, DC, AC+DC No of Outputs 3 standard 1 o
User Manual 2253i / 2253iX VOLT SURGE This test will produce voltage surge defined by Table 9-18. This test will be repeated three times with s in
User Manual 2253i / 2253iX 10. Option -704: MIL-STD 704 Rev D & E (MIL704 Mode) 10.1 General Option –704 includes a firmware implementation fo
User Manual 2253i / 2253iX 10.4 Available MIL-STD 704 Tests 10.4.1 STEADY STATE AC Mode: 1. Steady State Voltage an eon test 6. Frequency Modula
User Manual 2253i / 2253iX 10.5 Front Panel Operation MIL704 To perform a test from the keyboard, from the MENU screen, select the APPLICATIONS sc
User Manual 2253i / 2253iX 10.6 AC Test Mode Following sections cover testing in AC output mode. 10.6.1 Steady State Tests Scroll to the STEADY S
User Manual 2253i / 2253iX FREQUENCY This test will change the output frequency in the sequence shown in Table 10-2. SEQUENCE FREQUENCY TIME 40
User Manual 2253i / 2253iX WAVEFThis test will generate a 5% THD voltage distortion on the output voltage waveform. Using a clipped sine wave cause
User Manual 2253i / 2253iX • 420Hz for 4 sec. • 410Hz for 5 sec. • 61Hz for 0.5 sec. • 60.5Hz for 0.5 sec. ← key (backspace) will terminate the
User Manual 2253i / 2253iX 10.6.2 Emergency Test From the MIL704 main menu (Figure 10-2) scroll to the EMERGENCY entry using the up and down cursor
User Manual 2253i / 2253iX 10.6.3 Abn mor al Test From the MIL704 main menus as shown in Figure 10-2, scroll to the ABNORMAL AC entry using the up
User Manual 2253i / 2253iX Output Parameter Specification 6.5 Arms @ 115 V 19.5 Arms @ 115 V DC Mode High Voltage range 1.25 Adc @ 400 V 1.6
User Manual 2253i / 2253iX • 0V for 2 se• 100V for 8 seconds. conds. This test will change the output frequency in the sequence shown in Table 10
User Manual 2253i / 2253iX 10.7 DC Test Mode DC test mode requires DC mode and a steady state voltage setting of 28V DC or 270V DC. 10.7.1 Steady
User Manual 2253i / 2253iX 10.7.2 Transient Test From the MIL704 DC MENU scroll to the TRANSIENT DC entry using the up and down cursor keys. Press
User Manual 2253i / 2253iX 10.7.3 Abnormal Test From the MIL704 DC MENU scroll to the ABNORMAL DC entry using the up and down cursor keys. Press t
User Manual 2253i / 2253iX 10.7.4 Emergency Test From the MIL704 DC MENU scroll to the EMERGENCY DC entry using the up and down cursor keys (Figure
User Manual 2253i / 2253iX 11. Option –ABD: Airbus ABD0100.1.8 Test Use of this option requires the following: • Compact i/iX Series AC power sou
User Manual 2253i / 2253iX 12. Option –A350: Airbus ABD0100.1.8.1 Test Use of this option requires the following: • Compact i/iX Series AC power
User Manual 2253i / 2253iX 13. Option –AMD: Airbus AMD24 Test Use of this option requires the following: • Compact i/iX Series AC power source. •
User Manual 2253i / 2253iX 14. Option –B787: Boeing B787-0147 Test U• iXCGui Windows software. Provided on CI P/N CIC496 CD ROM or available for
User Manual 2253i / 2253iX 15. Option –WHM: Watt Hour Meter measurements The WHM measurement function can be accessed from the APPLICATIONS screen
User Manual 2253i / 2253iX Note: All output specifications apply below the Current / Voltage rating line shown in the V/I rating charts of section
User Manual 2253i / 2253iX 16. Err sor Me sages Any errors that occur during operation t pane terface will resu rro ssa he Lthe error message queue
User Manual 2253i / 2253iX Number Message String Cause Remedy -201 d while in local" ued but unit Put instrument in remote state before
User Manual 2253i / 2253iX Number Message String Cause Remedy lost" contents lost -315 "Configuration memory lost" settings l
User Manual 2253i / 2253iX Number Message String Cause Remedy influence (Load, voltage kickback, etc.) 802 "Current limit fault" t
User Manual 2253i / 2253iX California Instruments 154 Number Message String Cause Remedy 826 “Peak current prot trip” Peak currentexceeded.
User Manual 2253i / 2253iX California Instruments 155 17. Index 160... 27 704...
User Manual 2253i / 2253iX California Instruments 156 Mode output... 25 MODE ...
User Manual 2253i / 2253iX 2.1.2.1 Voltage versus Current Rating Charts – AC Mode – 230V AC input Figure 2-1: 2253i / iX Voltage / Current Rating
User Manual 2253i / 2253iX 2.1.2.2 Voltage versus Current Rating Charts – AC Mode – 115V AC input 2253i/iX Voltage Current Profile - AC Low Vrange
User Manual 2253i / 2253iX 2.1.2.3 Voltage versus Current Rating Charts – DC Mode 2253i/iX Voltage Current Profile - DC Low Vrange0.000.501.001.5
User Manual 2253i / 2253iX 2.1.3 Measurements Measurement specifications apply to single chassis AC sources. See notes for other models and configu
User Manual 2253i / 2253iX Refers to: Models: 2253i AC and DC Power Source 2253iX AC and DC Power Source/Analyzer Manual revision: E. Copyright © 2
User Manual 2253i / 2253iX 2.1.5 System Specification Controller Features Specification Trigger Input: External trigger source input. Requires TT
User Manual 2253i / 2253iX 2.1.6 Unit Protection Parameter Specification Input Over current: Input Circuit Breaker. This breaker protects the equ
User Manual 2253i / 2253iX 2.2 Mechanical Parameter Specification Dimensions: Height: 5.25 inches (13.3 cm) Depth: 23 inches (58.4 cm) Width:
User Manual 2253i / 2253iX 2.3 Environmental Parameter Specification Operating Temp: 0° to +40° C, full power. +32° to +104° F, full power. Storag
User Manual 2253i / 2253iX 2.4 Front Panel Controls, Indicators and Display Controls: Shuttle knob: The rotating knob may be used to adjust sett
User Manual 2253i / 2253iX 2.5 Special Features Controller Features Controller: Programmable controller front panel assembly. Mode Available sing
User Manual 2253i / 2253iX 2.6 Available Options – i Series Interface Options -GPIB GPIB Remote control interface. This option is not field insta
User Manual 2253i / 2253iX 2.7 Available Options – iX Series Interface Options -LAN Ethernet LAN interface connection. RJ45 connector. This optio
User Manual 2253i / 2253iX 20 KOhm. This option is mutually exclusive with the –LKS and –EXT option. -WHM Watt-hour measurement option. 2.8 EXT
User Manual 2253i / 2253iX 2.8.1 Typical Frequency Response – EXT Option Typical performance for EXT option models in the 300VAC range operating a
User Manual 2253i / 2253iX SAFETY SUMMARY This power source contains high voltage and current circuits, which are potentially lethal. Because of i
User Manual 2253i / 2253iX 2.9 LKM / LKS Options - Supplemental Specifications The Clock and Lock option enables two or more independent 22531iX p
User Manual 2253i / 2253iX 2.12 Supplemental Specifications Supplemental specifications are provided for reference only and are not guaranteed. Dat
User Manual 2253i / 2253iX 2.12.2 Remote Programming Output Parameter Specification Bus command response time: < 20 ms Ext. Trigger response t
User Manual 2253i / 2253iX 3. Unpacking and Installation 3.1 Unpacking Inspect the unit for any possible shipping damage immediately upon receipt
User Manual 2253i / 2253iX 3.3.2 Rackmount If the power source is to be mounted in a cabinet system, proper supports such as rack slides, brackets
User Manual 2253i / 2253iX 3.4 Rear Panel Connectors All connections to and from the power source are made at the rear panel. For the location of
User Manual 2253i / 2253iX 3.5 AC Input Wiring - INPUT AC input connecis located on thetions are to be made directly to the input terminal block.
User Manual 2253i / 2253iX 3.6 Output Connections 3.6.1 Output Wiring The output terminal block is located at the rear of the unit. Output connecti
User Manual 2253i / 2253iX 3.6.2 Output Terminal Block - OUTPUT Each chassis has a single AC output terminal strip. The output terminal strip must
User Manual 2253i / 2253iX 3.7 Connectors - Rear Panel A number of connectors are located along the top rear panel of the unit. A summary of avail
User Manual 2253i / 2253iX California Instruments 4
User Manual 2253i / 2253iX 3.7.1 AC Input Connector – INPUT – TB3 See section 3.5 for details on connecting AC input power. Connector type is KULK
User Manual 2253i / 2253iX 3.7.4 Auxiliary I/O Connector – J18 A high density D style, 15-pin I/O connector is located on the rear panel. Table 3-
User Manual 2253i / 2253iX 3.7.6 USB Interface – J16 A standard USB Series B device connector is located on the rear panel for remote control. A s
User Manual 2253i / 2253iX 3.7.8 GPIB Interface – J17 A standard IEEE488/ANSI MC1.1; 24 pin GPIB connector is located on the rear panel on al 2253
User Manual 2253i / 2253iX 3.7.9 LAN Interface – RJ45 – J15 An optional RJ45 Ethernet 10BaseT connector is located on the rear panel for remote co
User Manual 2253i / 2253iX 3.8 Basic Initial Functional Test CAUTION: Work carefully when performing these tests; hazardous voltages are presen
User Manual 2253i / 2253iX Model 115V on 150 V range 230V on 300 V range 2253i/iX, 3 phase mode 20 Ohm 80 Ohm 2253i/iX, 1 phase mode 6.67 Ohm
User Manual 2253i / 2253iX 3.9 Clock and Lock Mode (-LKM/-LKS Option) Clock and lock mode operation of two or more iX AC power sources is availabl
User Manual 2253i / 2253iX 3.9.1 Configuration settings Units configured with the –LKM option will show the Clock as INT (internal) and the mode as
User Manual 2253i / 2253iX 4. Front Panel Operation 4.1 Tour of the Front Panel Before operating the AC source using the front panel, it helps to
User Manual 2253i / 2253iX WARRANTY INFORMATION CALIFORNIA INSTRUMENTS CORPORATION warrants each instrument manufactured by them to be free from d
User Manual 2253i / 2253iX 4.1.3 Status Indicator Lights Five yellow LED status indicators are located on the left hand side of the keyboard/displ
User Manual 2253i / 2253iX 4.1.4 The Shuttle Knob CounterClockwiseClockwiseINCRDECR Figure 4-2: Shuttle Knob ting knob (“shuttle”). This shuttle k
User Manual 2253i / 2253iX 4.1.5 Menu Keys The i/iX Series is operated through a series of menus. These menus can be reached by using a number of m
User Manual 2253i / 2253iX 4.1.6 Cursor and Enter Keys The cursor keys are located on the right hand side of the numeric keypad and can be used to
User Manual 2253i / 2253iX 4.2 Me un Structure The next few pages show a map of the available menus in the i/iX Series. All menus can be reached b
User Manual 2253i / 2253iX 5. Memory test rPASSED" will esult. If all memory tests pass at power on, the message "MEMORY TEST appear.
User Manual 2253i / 2253iX 4.2.2 Top Level Menus The following op le can be accessed using the Menu ke RY ION t - vel menu choices y:ENT DESCR
User Manual 2253i / 2253iX 4.2.3 Menu Tree i/iX Series enu Structure EVEL 1 VEL 1 M L LEVEL 2 LE LEVEL 2 PROGRAM GISTERS VOLT
User Manual 2253i / 2253iX VOLT #12 120.0 LANGUAGE ON VSLEW #12 1e+06 ADVANCE ON FREQ #12 400.0 MODE
User Manual 2253i / 2253iX 4.2.4 PROGRAM Menu Figure 4-5: PROGRAM Menu hown in Figure 4-5. It can be reached in one of two ways: key, selecting t
User Manual 2253i / 2253iX Table of Contents 1. Introduction ...
User Manual 2253i / 2253iX ENTRY DESCRIPTION using the PHASE key first. CLIP LEVEL Sets the clip level for the CLIPPED sine wave in percent VTHD.
User Manual 2253i / 2253iX 4.2.5 CONTROL Menus Figure 4-6: CONTROL Menus Th CONTROL menu is sh 4selecting the CONTROL entry using the DThe
User Manual 2253i / 2253iX ENTRY DESCRIPTION STAN Power up in INT (internal) clock mode for stand-alone operation. This is the only mode for mo
User Manual 2253i / 2253iX ENTRY DESCRIPTION # OUTPUTS Selects SINGLE or THREE phase mode of operation. If the –MODE option is installed, this fi
User Manual 2253i / 2253iX 4.2.6 MEASUREMENTS Screens T a DS n regarding the output of the Source. This data acquisition system digitizes the volt
User Manual 2253i / 2253iX ENTRY DESCRIPTION CREST FACTOR This readout displays the ratio between peak current and rms current. MEASUREM NTS 3
User Manual 2253i / 2253iX 4.2.7 TRANSIENT Menu The transient menu is used to program and execse ences are defined as a se l list be exe
User Manual 2253i / 2253iX ENTRY DESCRIPTION higher than 200,000 are displayed as a floating point number (2E+05). The display has insufficient ch
User Manual 2253i / 2253iX ENTRY DESCRIPTION FUNC Step # Waveform selection. Available choices are Sinusoid, Square, Clipped or any of the user
User Manual 2253i / 2253iX 4.2.8 REGISTERS Menu The registers menu provides access to the non-volatile setup storage of the power source. A total
User Manual 2253i / 2253iX 6. Calibration...
User Manual 2253i / 2253iX 4.2.9 CONFIGURATION Menu The configuration menu may be used to con as the serial port, IEEE-488/GPIB address, LAN
User Manual 2253i / 2253iX ENTRY DESCRIPTION selected. See programming manual for details on use of either syntax. SYSTEM 2253iX This field is n
User Manual 2253i / 2253iX ENTRY DESCRIPTION GWAddress Gateway address setting. A default gateway is a node (a router) on a computer network that
User Manual 2253i / 2253iX 4.2.10 CALIBRATION Menus The measurement calibration menu can be used to perform routine calibration of the internal
User Manual 2253i / 2253iX 4.2.11 APPLICATIONS Menu Note that some of the application options list n may not be available on all i/iX models and ma
User Manual 2253i / 2253iX ENTRY DESCRIPTION LANGUAGE ON If –ABL option is installed, this field will indicate ON. N/A Default is N/A. All C
User Manual 2253i / 2253iX 4.2.13 Elapsed Time and Temperature Screen The Etime/Temp screen displays the elapsed time since the power source has
User Manual 2253i / 2253iX 4.2.14 LIMIT Menu , frequency and current range of the ins no user changeable fields. The The Limit menu displays
User Manual 2253i / 2253iX 4.3 Output Programming 4.3.1 Set the Output Output parameters are all set from the PROGRAM screen. 1. Use the MENU ke
User Manual 2253i / 2253iX 4.3.3 Change Output Values in SET Mode The SET mode of operation is a mode in which changes to output parameters made w
User Manual 2253i / 2253iX List of Figures Figure 1-1: Model 2253iX AC Source...
User Manual 2253i / 2253iX 4.4 Waveform Management The iX Series employs independent arbitrary waveform generators for each phase. This allows the
User Manual 2253i / 2253iX 4.4.3 RMS Amplitude Restrictions The out . If the AC source is in the 300 V range, the maximuwaveform is used however,
User Manual 2253i / 2253iX Fmaxh is thFmax is the upper frequency lime maximum allowable fundamental frequency that can be programmed it of the po
User Manual 2253i / 2253iX 4.5 Measurements Standard measurementmeasurements are spread across multiple screenss are always available through the M
User Manual 2253i / 2253iX frequency will not be reflected in the i/iX Series measurements. When using an external measurement reference, this may
User Manual 2253i / 2253iX 4.7.4 Pulse Transients Pulse transients let you program the output to a specified value for a predetermined amount of ti
User Manual 2253i / 2253iX 4.7.6 Pro ng list ents fro ont pgrammi transi m the fr anel The output transient system allows sequences of programmed
User Manual 2253i / 2253iX Figure 4-12: Sample Transient Output Sequence This output can be accomplished using the following transient list. Step
User Manual 2253i / 2253iX to be set to an integer number of periods. Over longlost due to timing skew between the waveform gene periods of time, p
User Manual 2253i / 2253iX 4.8 Setting the Power-on Initialization Values The power source is shipped with default factory settings when the unit
User Manual 2253i / 2253iX List of Tables Table 3-1: Output Terminal connections. ...
User Manual 2253i / 2253iX 4.9 Remote Inhibit Function The remote inhibit input used to disable the outpuavailable on the high density D-sub 15 co
User Manual 2253i / 2253iX 5. Principle of Operation 5.1 Overall Description The 2253i/iX models share a common architecture based on single 750VA
User Manual 2253i / 2253iX 5.6 Range/Relay Assembly The Range/Relay Assembly (P/N 6005-712-1) is located in the back of the chassis and contains th
User Manual 2253i / 2253iX CAUTION IN SECTIONS OF THIS . THIS EQUIPMENT GENERATES VOLTAGES UP TO 300 VAC AND 500 VDC ARE PRESENT IN CERTAPOW
User Manual 2253i / 2253iX 6. Calibration The Routine Calibration should be performed every 12 months. Non-routine Calibration is only required if
User Manual 2253i / 2253iX The shunt must be connected in series mOhm current shunt of sufficient powerwith the load. Connect the load to the outp
User Manual 2253i / 2253iX 6.4 Non-Routine Output Calibration The output calibration will not affect the accuracy of the output voltage when the A
User Manual 2253i / 2253iX 6.5 Non-Routine Output Offset and Gain Calibration WARNING: This requires the top cover to be removed and should be do
User Manual 2253i / 2253iX Figure 6-1: Internal adjustment locations. California Instruments 98
User Manual 2253i / 2253iX 7. Service 7.1 Cleaning The exterior of the power source may be cleaned with a cloth dampened with a mild detergent and